Skip Navigation

Europace 2005 7(3):255-265; doi:10.1016/j.eupc.2005.01.007
This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (3)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Biffi, M.
Right arrow Articles by Branzi, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Biffi, M.
Right arrow Articles by Branzi, A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© 2005 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.

Pacing with capture verification in candidates for resynchronisation therapy: A feasibility study

Mauro Biffia,*, Giuseppe Boriania, Matteo Bertinia, Paolo Silvestrib, Cristian Martignania and Angelo Branzia

aInstitute of Cardiology, University of Bologna Via Massarenti 9, 40138 Bologna, Italy; bSt. Jude Medical Italy

BACKGROUND: Devices for cardiac resynchronisation therapy (CRT) deliver energy into 3 output channels. Such a burden can significantly reduce device longevity. AutocaptureTM has been shown to improve pacemaker longevity and safety of right ventricular pacing in clinical studies. The aim of this study was to investigate the application of AutocaptureTM during biventricular pacing (BIV) to decrease the energy cost of CRT.

METHODS: During implantation of BIV devices, an acute study was performed to test the hypothesis that the evoked response (ER) elicited by each delivered stimulus is correctly detected and measured either on the right ventricular (RV) channel during BIV pacing with the left ventricular (LV) channel pacing first, or in the LV channel with the RV channel pacing first. A reliable measurement of ER is the critical requirement for the correct performance of AutocaptureTM.

RESULTS: ER amplitude in the right ventricle during BIV pacing was not significantly decreased compared with RV pacing in the VVI mode (16.36 ± 5.27 mV vs 17.09 ± 6.12 mV). ER amplitude in the left ventricle during BIV pacing was not significantly decreased compared with LV pacing in the VVI mode (12.4 ± 8.95 mV vs 12.25 ± 8.97 mV). Three patients in atrial fibrillation had a DDDR pacemaker with the LV lead connected to the atrial port, and received BIV pacing with AutocaptureTM turned on in the RV channel. AutocaptureTM performance in the long term, as assessed by the trend of RV threshold over 20 ± 8 months, showed that LV depolarisation was never sensed as an ER on the RV channel.

CONCLUSIONS: Our observations support the feasibility and safety of capture verification during BIV pacing on the ventricular channel paced secondly, which could increase the longevity of CRT devices, and decrease the costs of this new therapy for heart failure patients.

Key Words: heart failure, resynchronisation therapy, autocapture, evoked response


*Corresponding author. Tel.: +39 051 6363531; fax: +39 051 344859. E-mail address: mbiffi{at}orsola-malpighi.med.unibo.it (M. Biffi).


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Eur Heart J SupplHome page
M. Biffi, J. Sperzel, C. Martignani, A. Branzi, and G. Boriani
Evolution of pacing for bradycardia: Autocapture
Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I23 - I32.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
M. Biffi, M. Bertini, M. Ziacchi, and G. Boriani
Left ventricular pacing by automatic capture verification
Europace, December 1, 2007; 9(12): 1177 - 1181.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.