Arrhythmia detection by dual-chamber implantable cardioverter defibrillators
A review of current algorithms
aNancy University Hospital Nancy, France; bELA Medical, Le Plessis-Robinson France
This article reviews the arrhythmia detection criteria currently available in dual-chamber implantable cardioverter defibrillators (ICDs), and describes the implementation and performance of various detection algorithms. Nearly all criteria implemented in single-chamber devices appear to have been included in dual-chamber ICDs. However, two different strategies can be distinguished: the first adds dual-chamber inhibition criteria to a single-chamber detection configuration; the second is a new approach entirely based on a dual-chamber detection scheme. Despite widely available clinical data, an analysis of the implemented detection indexes, arrhythmia characteristics (induced vs spontaneous, minimum duration), device programming (detection rate, programme maintenance and tuning during follow-up), and different storage capabilities among various ICD models, leaves the results of these studies ultimately ambiguous. New algorithms are under study, but only protocols using a single set of arrhythmias and the same programming for all devices may allow relevant comparisons of the performances of detection algorithms. Furthermore, a criterion is required to distinguish reliably between haemodynamically stable and unstable tachycardias, not simply based on rate, but including the underlying cardiac function.
Key Words: implantable cardioverter defibrillator, dual-chamber, arrhythmia detection algorithm
*Corresponding author. Service Cardiologie, Hôpital Brabois, Rue du Morvan, 54500 Vandoeuvre-les-Nancy, France. Tel.: +33-383-153244; fax: +33-383-153856. E-mail address: E-mail address: e.aliot{at}chu-nancy.fr (E. Aliot).
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
S. Tzeis, G. Andrikopoulos, C. Kolb, and P. E. Vardas Tools and strategies for the reduction of inappropriate implantable cardioverter defibrillator shocks Europace, November 1, 2008; 10(11): 1256 - 1265. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Boriani, E. Occhetta, S. Cesario, S. Grossi, M. Marconi, G. Speca, P. Silvestri, M. Biffi, M. Bortnik, C. Martignani, et al. Contribution of morphology discrimination algorithm for improving rhythm discrimination in slow and fast ventricular tachycardia zones in dual-chamber implantable cardioverter-defibrillators Europace, August 1, 2008; 10(8): 918 - 925. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Boriani, R. Ricci, T. Toselli, R. Ferrari, A. Branzi, and M. Santini Implantable cardioverter defibrillators: from evidence of trials to clinical practice Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I66 - I73. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. H. Schoenfeld Contemporary Pacemaker and Defibrillator Device Therapy: Challenges Confronting the General Cardiologist Circulation, February 6, 2007; 115(5): 638 - 653. [Full Text] [PDF] |
||||