Skip Navigation

Europace 2001 3(4):324-331; doi:10.1053/eupc.2001.0188
© 2001 by European Society of Cardiology
This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (4)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Santini, M.
Right arrow Articles by Caporicci, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Santini, M.
Right arrow Articles by Caporicci, D.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?


PACING

Immediate and long-term atrial sensing stability in single-lead VDD pacing depends on right atrial dimensions

M. Santini, R. Ricci, C. Pignalberi, A. Auriti, M. Pepe, R. Assale1 and D. Caporicci2

Department of Cardiology, S. Filippo Neri Hospital Rome; 1Aosta Hospital Aosta; 2Macerata Hospital Macerata-Italy

Abstract

AIMS: Single-pass VDD is a physiological stimulation mode which requires a single-lead with a floating dipole to detect the atrial signal. We investigated the impact of right heart dimensions on immediate and long-term atrial sensing stability in VDD systems to draw guidelines for optimal atrial sensitivity programming.

METHODS: Forty-one patients (23 males, mean age 73±11 years) with II or III degree AV block and normal sinus node function received a Thera VDD Medtronic 8948 pulse generator with Medtronic 5032 lead. Atrial sensing was evaluated at pre-discharge and during 12-months follow-up in the supine and upright positions, during normal and forced breathing. Atrial sensing variability, quantified by a D value (mV), which represents the difference between the maximum and the minimum atrial electrogram amplitude obtained during the various activities, was analyzed and related to right heart dimensions measured at echocardiography.

RESULTS: Long-term appropriate atrial sensing was obtained in 40/41 patients. P wave mean amplitude at pre-discharge, during every specific activity, was predictive of 12-month values. No atrial oversensing was observed. According to the right atrial dimension (RAD), patients with RAD >38 mm had a sensing variability significantly higher than those with RAD ≤34 mm (D=1±1 Vs D=0·4±0·5;P< 0·05). A linear relationship between D and RAD dimensions was observed (r=0·47;P< 0·005). According to the distance from the roof of the right atrium to the right ventricular apex (RAVD), patients with RAVD <93 mm had D=1·1±1, while patients with RAVD >93 mm had D=0·5±0·4;P< 0·05); a inverse correlation between RAVD and D was highlighted (r=–0·43;P< 0·01). Using a cluster analysis, the combination of RAD and RAVD identified a subgroup (RAD >30 mm and RAVD <80 mm) at high risk of loss of AV synchrony with 67% sensitivity and 80% specificity. Nevertheless, apart from heart dimensions, no patient showed an atrial signal amplitude lower than 0·3 mV.

CONCLUSIONS: According to our data, in VDD single-lead systems the amplitude of the atrial signal is stable over time in every physical activity. High RAD and low RAVD values may select patients with poor atrial sensing stability. Anyway, taking into account that no atrial oversensing was observed, atrial sensitivity setting at the highest value should be recommended.

Key Words: VDD pacing mode, physiological pacing, echocardiography, right heart dimensions, atrial sensing


Correspondence: Prof Massimo Santini, Department of Cardiology-Ospedale S. Filippo Neri Via Martinotti 20-00135 Roma, Italy. E-mail: m.santini{at}rmnet.it


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.