The use of permanent pacemakers in the detection of cardiac arrhythmias
Department of Cardiology, Freeman Hospital Newcastle upon Tyne, U.K.
AIMS: To compare pacemaker telemetry with simultaneous Holter recordings in the diagnosis of atrial fibrillation and to evaluate the STOP-AF study telemetry criteria for the presence of atrial fibrillation.
METHODS AND RESULTS: 18 consecutive patients enrolled in the STOP-AF study had simultaneous 24 h Holter recordings and down-loaded pacemaker telemetry. There was good agreement on heart rate, but the STOP-AF pacemaker criteria achieved only 57% sensitivity with 64% specificity for the presence of atrial fibrillation on Holter recording over 24 h. False-positives appeared to result from far-field sensing while false-negatives occurred with very short episodes of atrial fibrillation. The pacemaker's antitachycardia responses were not specific for the presence of atrial fibrillation.
CONCLUSION: Pacemaker telemetry is a potentially important source of data on cardiac arrhythmias. Further studies are required to define the limitations of these data in specific devices before they can be interpreted with confidence.
Key Words: Pacemaker, atrial fibrillation, Holter monitoring, telemetry
Correspondence: Dr C. J. Plummer, c/o Dr McComb, Department of Cardiology, Freeman Hospital, Newcastle upon Tyne NE7 7DN, U.K. E-mail: C.J.Plummer{at}ncl.ac.uk
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
B. K. Kantharia, R. A. Freedman, D. Hoekenga, G. Tomassoni, S. Worley, R. Sorrentino, D. Steinhaus, J. M. Wolkowicz, Z. A. Syed, and and the AOP Study Investigators Increased base rate of atrial pacing for prevention of atrial fibrillation after implantation of a dual-chamber pacemaker: insights from the Atrial Overdrive Pacing Study Europace, November 1, 2007; 9(11): 1024 - 1030. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Schuchert, S. Lepage, J.J. Ostrander, R.J. Bos, M. Gwechenberger, A. Nicholls, and G. Ringwald Automatic analysis of pacemaker diagnostic data in the identification of atrial tachyarrhythmias in patients with no prior history of them Europace, January 1, 2005; 7(3): 242 - 247. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Kale and D. H. Bennett Pacemaker prevention therapies for the control of drug-refractory paroxysmal atrial fibrillation Europace, January 1, 2003; 5(2): 123 - 131. [Abstract] [PDF] |
||||