AIMS: Assessment of complications following implantation of transvenous ventricular electrodes to pace the left ventricle.
METHODS AND RESULTS: Twenty-eight patients with severe cardiac failure and left bundle branch block were prospectively followed for adverse effects of implantation of a left ventricular transvenous pacing system. Immediate followup was associated with loss of left ventricular pacing in nine patients (32%). This was due to lead dislodgement in four cases (corrected by re-operation in three of these cases), and due to increased threshold in five cases (corrected by programming a higher pacing amplitude in all five cases, but with intermittent diaphragmatic contraction in one case). After 1 month, one patient died, one patient with severe coronary heart disease suffered a myocardial infarction, and left ventricular pacing was lost in two patients. Pericardial effusion, new significant ventricular arrhythmias or other adverse effects were not observed. After a mean follow-up of 16±9·2 months, pacing leads remained stable and no late complications related to the transvenous left ventricular epicardial pacing were observed.
CONCLUSION: Placement of a permanent lead in a tributary of the coronary sinus is feasible without serious adverse effects during the first month. The only frequent adverse event was lead dislodgement; a finding which emphasizes the need for development of specially designed leads for this application.
Key Words: Heart failure, spacing, adverse events
Correspondence: Jean-Jacques Blanc, Department of Cardiology, Brest University Hospital, France. E-mail: Jean-Jacques.Blanc{at}univ.brest.fr
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
G. Berberian, T. A. Quinn, S. E. Cabreriza, J.-E. S. Kenny, C. A. Garofalo, A. D. Weinberg, and H. M. Spotnitz Left ventricular pacing site and timing optimization during biventricular pacing using a multielectrode patch in pigs J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 574 - 578. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Bulava and J. Lukl Single-centre experience with coronary sinus lead stability and long-term pacing parameters Europace, July 1, 2007; 9(7): 523 - 527. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Berberian, S. E. Cabreriza, T. A. Quinn, C. A. Garofalo, and H. M. Spotnitz Left Ventricular Pacing Site-Timing Optimization During Biventricular Pacing Using a Multi-Electrode Patch Ann. Thorac. Surg., December 1, 2006; 82(6): 2292 - 2294. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Mair, J. Sachweh, B. Meuris, G. Nollert, M. Schmoeckel, A. Schuetz, B. Reichart, and S. Daebritz Surgical epicardial left ventricular lead versus coronary sinus lead placement in biventricular pacing Eur. J. Cardiothorac. Surg., February 1, 2005; 27(2): 235 - 242. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. J. Drew, R. M. Califf, M. Funk, E. S. Kaufman, M. W. Krucoff, M. M. Laks, P. W. Macfarlane, C. Sommargren, S. Swiryn, and G. F. Van Hare Practice Standards for Electrocardiographic Monitoring in Hospital Settings: An American Heart Association Scientific Statement From the Councils on Cardiovascular Nursing, Clinical Cardiology, and Cardiovascular Disease in the Young: Endorsed by the International Society of Computerized Electrocardiology and the American Association of Critical-Care Nurses Circulation, October 26, 2004; 110(17): 2721 - 2746. [Abstract] [Full Text] [PDF] |
||||
![]() |
A.L.A.J. Dekker, B. Phelps, B. Dijkman, T. van Der Nagel, F.H. van Der Veen, G.G. Geskes, and J.G. Maessen Epicardial left ventricular lead placement for cardiac resynchronization therapy: Optimal pace site selection with pressure-volume loops J. Thorac. Cardiovasc. Surg., June 1, 2004; 127(6): 1641 - 1647. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.G. Maessen, B. Phelps, A.L.A.J. Dekker, and B. Dijkman Minimal invasive epicardial lead implantation: optimizing cardiac resynchronization with a new mapping device for epicardial lead placement Eur. J. Cardiothorac. Surg., May 1, 2004; 25(5): 894 - 896. [Abstract] [Full Text] [PDF] |
||||
![]() |
V.ér. Valls-Bertault, M. Fatemi, M. Gilard, P. Y. Pennec, Y. Etienne, and J.-J. Blanc Assessment of upgrading to biventricular pacing in patients with right ventricular pacing and congestive heart failure after atrioventricular junctional ablation for chronic atrial fibrillation Europace, January 1, 2004; 6(5): 438 - 443. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. T. Abraham, W. G. Fisher, A. L. Smith, D. B. Delurgio, A. R. Leon, E. Loh, D. Z. Kocovic, M. Packer, A. L. Clavell, D. L. Hayes, et al. Cardiac Resynchronization in Chronic Heart Failure N. Engl. J. Med., June 13, 2002; 346(24): 1845 - 1853. [Abstract] [Full Text] [PDF] |
||||