Skip Navigation

Europace 2000 2(3):240-244; doi:10.1053/eupc.2000.0103
© 2000 by European Society of Cardiology
This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (1)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Cornacchia, D.
Right arrow Articles by de Seta, F.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Cornacchia, D.
Right arrow Articles by de Seta, F.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Latest generation of unipolar and bipolar steroid eluting leads

Long-term comparison of electrical performance in atrium and ventricles

D. Cornacchia1, M. Fabbri1, A. Puglisi2, P. Moracchini3, M. Bernasconi4, M. Nastasi5, C. Menozzi6, G. Mascioli7, T. Marotta8 and F. de Seta8

1Cardiology Dept, Degli Infermi Hospital Feanza (RA), Italy; 2Cardiology Dept, Fatebenefratelli Hospital Rome, Italy; 3Cardiology Dept, S. Agostino Hospital Modena, Italy; 4Cardiology Dept, S. Carlo Borromeo Hospital Milan, Italy; 5Cardiology Dept, Casa di Cura Villabianca Bari, Italy; 6Cardiology Dept, S. Maria Nuova Hospital Reggio Emilia, Italy; 7Cardiology Dept, Della Val Camonica Hospital Esine, Italy; 8Medtronic Italy Italy

One hundred and seventy-three patients, mean age 74 years permanently paced with 123 atrial (53 unipolar, 70 bipolar) and 143 ventricular (73 unipolar, 70 bipolar) pacing leads were included in this study. The pacing leads were recent generation low surface area steroid eluting leads from one manufacturer: leads with silicone and polyurethane insulation were studied, and they were combined with generations of one pacemaker family from the same manufacturer permitting identical measurements to be made over a follow-up of 2 years. Pacing threshold was measured using pulse duration at a fixed voltage of 1·5 V: peak to peak P and R wave amplitude and pacing impedance at 2·5 V and 0·5 ms were all measured using the manufacturer's standard programmer. Although many significant differences, in the parameters measures, existed between atrium and ventricle and unipolar and bipolar configurations, none was felt to be of clinical significance. These data permit the physician to choose the lead type with regard to sensing performance and long-term lead integrity.

Key Words: Unipolar, bipolar atrial and ventricular pacing leads, steroid elution, stimulation threshold, sensing


Correspondence: Dr Daniele Cornacchia, Via Orto Caporal, 5, 48018, Faenza (RA) Italy.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.