Skip Navigation


Europace Advance Access originally published online on February 26, 2009
Europace 2009 11(5):565-570; doi:10.1093/europace/eup040
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
11/5/565    most recent
eup040v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in Europace
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Linhart, M.
Right arrow Articles by Lewalter, T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Linhart, M.
Right arrow Articles by Lewalter, T.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2009. For permissions please email: journals.permissions@oxfordjournals.org


Atrial Fibrillation Registry

In vitro comparison of platinum–iridium and gold tip electrodes: lesion depth in 4 mm, 8 mm, and irrigated-tip radiofrequency ablation catheters

Markus Linhart1,* {dagger}, Hanke Mollnau2 {dagger}, Alexander Bitzen1, Sabine Wurtz3, Jan W. Schrickel1, René Andrié1, Florian Stöckigt1, Christian Weiß2, Georg Nickenig1, Lars M. Lickfett1 and Thorsten Lewalter1

1 Department of Medicine-Cardiology, University of Bonn, Sigmund-Freud-Str. 25, 53105 Bonn, Germany; 2 Department of Medicine II, Johannes Gutenberg–University, Langenbeckstraße 1, 55131 Mainz, Germany; 3 Biotronik GmbH & Co. KG, Berlin, Woermannkehre 1, 12359 Berlin, Germany

Aims: We compared a newly developed irrigated gold tip electrode ablation catheter and a gold tip 4 and 8 mm catheter with the corresponding platinum–iridium (Pt) tip catheters in an in vitro setting.

Methods and results: In a flow chamber simulating physiological flow conditions, radiofrequency catheter ablation was performed on tissue samples of porcine endomyocardium and liver. Lesion depth, energy and temperature delivery, and popping frequency were determined. Two hundred and fifty-three ablations were conducted. Four and eight millimetre, gold tip electrode catheters produced significantly deeper lesions compared with the Pt tip electrode (liver 4 mm: 4.67 ± 1.7 vs. 2.9 ± 1.0 mm, P < 0.0001; endomyocardium 4 mm: 3.88 ± 1.1 vs. 2.81 ± 0.7 mm, P < 0.001; liver 8 mm: 3.98 ± 1.0 vs. 2.03 ± 1.1 mm, P < 0.001; endomyocardium 8 mm: 4.00 ± 0.9 vs. 3.39 ± 0.8 mm, P < 0.001) and correlated with the amount of energy delivery. Popping frequency was significantly higher in gold tip electrodes. In irrigated tip electrodes, there was no difference in the lesion depth comparing gold with Pt (liver: 5.18 ± 0.7 vs. 5.01 ± 0.7 mm, P = ns; endomyocardium: 4.89 ± 0.7 vs. 4.78 ± 0.8 mm, P = ns). There was a trend towards less popping in the gold tip electrode.

Conclusion: Both 4 and 8 mm not-irrigated gold tip catheters produced deeper lesions than the corresponding Pt tip catheter. In irrigated tip catheters, gold and Pt tip material did not show differences in the lesion depth.

Key Words: Radiofrequency ablation, Gold tip electrode, Platinum–iridium tip electrode, Irrigated tip catheter, Cooled tip catheter, Lesion depth


* Corresponding author. Tel: +49 228 287 5507, Fax: +49 228 287 1559, Email: markus.linhart{at}ukb.uni-bonn.de

{dagger} These authors contributed equally to this work.

Manuscript submitted 18 October 2008. Accepted after revision 12 January 2009.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?

Related articles in Europace:

Should we be considering gold-tipped ablation electrodes?
Hugh Calkins
Europace 2009 11: 541. [Full Text]  



This article has been cited by other articles:


Home page
EuropaceHome page
H. Calkins
Should we be considering gold-tipped ablation electrodes?
Europace, May 1, 2009; 11(5): 541 - 541.
[Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.