Skip Navigation


Europace Advance Access originally published online on August 7, 2009
Europace 2009 11(11):1505-1509; doi:10.1093/europace/eup222
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
11/11/1505    most recent
eup222v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in Europace
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (1)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Calvagna, G. M.
Right arrow Articles by Valsecchi, S.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Calvagna, G. M.
Right arrow Articles by Valsecchi, S.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2009. For permissions please email: journals.permissions@oxfordjournals.org.


Leads and Lead Extraction

Single-operator experience with a mechanical approach for removal of pacing and implantable defibrillator leads

Giuseppe M. Calvagna1,*, Rosario Evola1, Giuseppe Scardace2 and Sergio Valsecchi2

1 Division of Cardiology, San Vincenzo Hospital, Contrada Sirina, Taormina, 38068 Messina, Italy; 2 Medtronic Italia, Rome, Italy

Aims: Recently, a mechanical single-sheath technique with a multiple venous entry-site approach for the removal of pacemaker and implantable defibrillator leads was reported to have a high success rate and few complications. In our institution, this technique of lead removal has been used since 2002. In this paper, we report our experience, with the aim of evaluating the effectiveness and safety of the proposed procedure.

Methods and results: This study is a retrospective analysis of the case records of all patients referred to our institution for transvenous lead extraction, according to class I or II Heart Rhythm Society indications. Over 7 years, 300 consecutive patients underwent procedures for transvenous removal of 518 leads. The most frequent indication for extraction was infection (74%). Complete removal of 502 (96.9%) leads and partial removal of 10 leads (1.9%) were achieved. Six leads (1.2%) could not be removed. All defibrillation coils and coronary sinus leads were successfully removed. There were no procedure-related deaths but only one major complication (0.3%).

Conclusion: Our experience shows that the proposed mechanical technique is very effective and associated with few serious complications, thus confirming previous findings. This approach may be reproduced in other settings with very satisfactory results.

Key Words: Pacemaker, Implantable defibrillator, Lead, Extraction, Infection


* Corresponding author. Tel: +39 0942579214, Email: gicalvagna{at}tiscali.it

Manuscript submitted 27 March 2009. Accepted after revision 20 July 2009.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?

Related articles in Europace:

Cardiac implantable electronic device treatment: taking care of complications
Charles Kennergren
Europace 2009 11: 1419-1420. [Full Text]  



This article has been cited by other articles:


Home page
EuropaceHome page
C. Kennergren
Cardiac implantable electronic device treatment: taking care of complications
Europace, November 1, 2009; 11(11): 1419 - 1420.
[Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.