Skip Navigation


Europace Advance Access originally published online on August 13, 2008
Europace 2008 10(10):1152-1156; doi:10.1093/europace/eun214
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
10/10/1152    most recent
eun214v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Tomaske, M.
Right arrow Articles by Bauersfeld, U.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tomaske, M.
Right arrow Articles by Bauersfeld, U.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org


Implantable cardioverter-defibrillators

Epicardial and pleural lead ICD systems in children and adolescents maintain functionality over 5 years

Maren Tomaske1,*, René Prêtre2, Mariette Rahn2 and Urs Bauersfeld1

1 Division of Paediatric Cardiology, University Children's Hospital, Steinwiesstrasse 75, CH-8032 Zurich, Switzerland; 2 Division of Congenital Cardiovascular Surgery, University Children's Hospital, Zurich, Switzerland

Aims: The optimal implantable cardioverter defibrillator (ICD) system implant technique has not yet been defined in young patients and those with congenital heart disease (CHD). We describe our 5-year experience with epicardial pacing/sensing leads secured on the left cardiac chambers and a pleural defibrillation lead insertion along the third intercostal space.

Methods and results: Implantable cardioverter defibrillator systems were implanted in 15 children and adolescents (age: 2.9–20.0 years) for primary (n = 11) or secondary (n = 4) prevention. Underlying CHD were hypertrophic (n = 10) or dilative cardiomyopathies (n = 2), primary electrical diseases (n = 2), and transposition of the great arteries (n = 1). Devices were placed in the rectus sheath (n = 5), or within the diaphragm (n = 10). Median defibrillation threshold at implant was 15 J (range: 10–25). During 5 years of follow-up (median: 22 months), nine appropriate and two inappropriate ICD discharges occurred. Four system revisions were required due to device recall, pleural lead dislodgement, epicardial lead fracture, and insulation break. Twelve months after the implantation, defibrillation threshold testing demonstrated stable thresholds of ≤20 J in five patients.

Conclusion: Our 5-year experience demonstrates the efficacy of epicardial and pleural lead ICD systems. Inappropriate shocks and lead failures are observed as in other ICD systems. It represents an alternative implant technique for young and active patients and those without venous access.

Key Words: Implantable cardioverter defibrillator, Congenital heart disease, Implantation technique, Paediatric


* Corresponding author. Tel: +41 44 2667519; fax: +41 44 2667981.E-mail address: maren.tomaske{at}kispi.uzh.ch

Manuscript submitted 25 May 2008. Accepted after revision 25 July 2008.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.