Skip Navigation

Europace 1999 1(2):135-139; doi:10.1053/eupc.1998.0028
© 1999 by European Society of Cardiology
This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (6)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Neuzner, J.
Right arrow Articles by Haberbosch, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Neuzner, J.
Right arrow Articles by Haberbosch, W.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Bipolar atrial sensing thresholds in sinus rhythm and atrial tachyarrhythmias

A comparative analysis in patients with DDDR pacemakers

J. Neuzner, J. Sperzel, H. F. Pitschner, T. Schwarz, W. Ehrlich, B. Schulte and W. Haberbosch

Department of Cardiology, Kerckhoff Clinic Bad Nauheim, Germany

Automatic mode switching (AMS) function in dual chamber pacemakers depends on adequate detection of atrial tachyarrhythmias. There are few data on showing how intra-operative atrial signal amplititude during sinus rhythm can predict atrial tachyarrhythmias after pacemaker implantation. In 43 patients undergoing DDDR pacemaker implantation and atrioventricular nodal ablation for the treatment of drug-refractory paroxysmal atrial fibrillation, atrial sensing thresholds during sinus rhythm and during induced atrial tachyarrhythmias (24–48 h after device implantation) were analysed. Five different DDDR pacemaker systems were implanted (Chorus 7034|Pt, Ela Medical n=13; Meta DDDR 1254|Pt, Telectronics Pacing Systems n=12; Vigor DR 1230|Pt, Guidant n=6; Trilogy DR 2364|Pt, Pacesetter, n=2; Kappa DR 401|Pt, Medtronic USA n=10). Every patient received a steroid-eluting, screw-in, bipolar atrial lead (Medtronic, Capsure-Fix 4068|Pt). The mean P wave amplitude during implantation was 3·91±1·14 mV. The mean atrial sensing threshold during sinus rhythm and during all modes of induced atrial tachy-arrhythmias was 3·35±1·0 mV, and 1·52±0·92 mV, respectively (P<0·001). Atrial fibrillation was induced in 36 patients. The mean sensing threshold during sinus rhythm in this patient group was 3·39±1·01 mV, the mean sensing threshold during atrial fibrillation was 1·27±0·56 mV, reflecting a 63% reduction of sensing threshold compared with sinus rhythm (P<0·001). Atrial flutter was induced in seven patients. The mean sensing threshold during sinus rhythm was 2·92±1·19 mV, the mean sensing threshold during atrial flutter was 2·79±1·26 mV, reflecting a reduction of 5% (ns) compared with sinus rhythm. Atrial sensing thresholds during sinus rhythm were significantly correlated with sensing thresholds during atrial tachyarrhythmias (r=0·44;P<0·002), but there were significant variations in intra-individual results. The reduction of atrial sensing thresholds between sinus rhythm and induced atrial tachyarrhythmias ranged from 30% to 82%.

CONCLUSION: Bipolar atrial sensing thresholds during sinus rhythm are correlated with sensing thresholds during atrial tachyarrhythmias, but there is a large degree of variance in individual patients. A 4:1 to 5:1 atrial sensing safety margin based on sensing threshold during sinus rhythm is a predictor for adequate postoperative detection of atrial tachyarrhythmias and the function of AMS devices.

Key Words: Bipolar atrial sensing thresholds, automatic mode switch, atrial tachyarrhythmias, dual chamber pacing


Correspondence: Jörg Neuzner, Department of Cardiology, Kerckhoff Clinic, Beneke Strasse 2–8, 61231 Bad Nauheim, Germany.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
EuropaceHome page
B. Schmidt, S. Asbach, O. Schweika, M. Zehender, C. Bode, and T. S. Faber
Atrial fibrillation reduces the atrial impedance amplitude during cardiac cycle: a novel detection algorithm to improve recognition of atrial fibrillation in pacemaker patients
Europace, September 1, 2007; 9(9): 812 - 816.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
A. Schuchert, M. A. Aydin, C. Israel, G. Gaby, and V. Paul
Atrial pacing and sensing characteristics in heart failure patients undergoing cardiac resynchronization therapy
Europace, January 1, 2005; 7(2): 165 - 169.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.